![]()
![]()
|
Light meter for visible and UV light for PC RS232 / USB series B103 |
|
||||||||||||||||||||||
|
Description:
The sensor measures light
intensity in micro Watt / square cm. UVL = UV light intensity RED = Red light GRE = Green light BLU = Blue light
|
||||||||||||||||||||||
|
||||||||||||||||||||||
|
User Instructions for installation in a hazardous area Class 1 (Group A, B, C and D) |
||||||||||||||||||||||
|
Personalize: Like any analytical instrument there are many other things you can
do with it. The output from your measurement
shows in the PC and instead of using the calculated data’s from the
sensor you can create your own calibration by using the formula above or
other manipulations.
Calibration: Instant analysis calibrates the light intensity in the
microprocessor from the sensor. To recalibrate simply use your own
calibration method = light intensity and correct the incoming data’s
from the sensor. You always can use the values from the sensor as default.
You can calculate the new values or
put the formula in your PC software. Calibrated
to 10 micro watt / square cm for max output.
|
||||||||||||||||||||||
|
Figure 1: Spectral Intensity of UV and visible light calibrated to 10 micro watt / square cm for max output. Figure 2: Same as figure 1 with the addition of a typical light bulb and its output.
Example to show total light energy (figure 1): A halogen light with 3000 Lx is shown into the sensors aperture and the following calculations are done: a) total energy: convert 3000 Lx to Watts - 0.0015 × 3000 W = 5 Watts b) This energy has to be seen in a sphere of 1 meter radius, the surface area is 12.5 cm2 c) The sensors aperture is 10 mm2 d) The light energy going in to the sensor in one meter is: (5 watts × 10 mm2) ÷ (10 cm2) = 5 micro Watts
Example to show light measured with a sensor (figure 2): The total amount of the energy of the light bulb includes a part of the infrared light. Light energy seen over the spectra is measured in the light sensor. Light energy for each color is the following: The integral from the sensitivity of the sensor in its spectra x part of the light energy from the spectra going in the sensor. In this example the sensor shows the following outputs: UV - 1.33 micro Watt / cm2 Blue - 1.63 micro Watt / cm2 Green - 2.38 micro Watt / cm2 Red - 3.12 micro Watt / cm2 |
||||||||||||||||||||||
|
||||||||||||||||||||||
|
||||||||||||||||||||||